Varian 21 Series

Varian Clinac 2100C - Dual Energy Linear Accelerator
The Varian CLINAC 2100C is a high energy linear accelerator with a Klystron Modulator and a digital control system. Although upgradeable to MLC, supply of pre-owned MLC equipped units is very limited.

Specification
  2100C 2100C/D 21EX 2300C/D 23EX
Low Photon Energy MV 4 or 6 4 or 6 4 or 6 6 4 or 6
Dose Rate MU/min 50-250 100-600 (4MV) 50-250; (6MV) 100-600 100-600 (4MV) 50-250; (6MV) 100-600
High Photon Energy MV up to 23 up to 23 up to 23 up to 23 up to 25
Dose Rate MU/min 80-400 100-600 100-600 100-600 100-600
Electron Energy MeV 4 to 20 6 to 20 6 to 20 4 to 22 4-22Mev
Electron Dose Rate MU/min 80 to 400 100 to 1000 100-1000 100-1000 100-1000
Available with MLC Optional Optional Optional Optional Optional
Minimum Room Dimensions:          
Depth 24' 24' 24' 24' 24'
Width 20' 20' 20' 20' 20'
Height 10' 10' 10' 10' 10'
Power Requirements KVA 35 35 35 36 36
Water Requirements Gal/min 9 9 9 10 10